Description
- Four operating modes : CMV, A/CMV, SIMV, SPONT
- Lightweight, durable and simple to use
- Ventilation volume and ventilation frequency can be adjusted
in succession and separately - LCD graphical display to indicate airway pressure, alarm limits,
tidal volume, mode - Automatic pressure relief valve for airway pressure more
than 6KPa - Air-oxygen blender prolongs the operating time during
long-time transportation - Audiovisual low input pressure alarm and over input pressure
protection
SPECIFICATION
VENTILATOR |
RANGE |
Tidal Volume (at |
50~1200 ml |
Flow (at A/C, |
Low, Med, High |
Breath Rate |
5-60 bpm, 1~12 |
2% |
45%~90% |
Spont Flow |
0~5 L/Min |
Sigh |
1 per 0~100 |
Inspiratory time |
0.5~3 s |
Trigger Level |
-10~0 hpa |
PIP |
0~60 hpa |
Silence |
60s |
ALARMS
AND INDICATORS
High Pressure |
20~60hpa (to be |
Low Pressure |
2~20hpa (to be |
Alarm Indicator |
Flashing Red |
Over Pressure |
<60 hpa |
Main Circuit |
-2~10kPa |
ELECTRICAL SPECIFICATION
INPUT POWER SUPPLY
- DC Voltage : 12 V
- Power : 10 VA
CHARGER
- AC Voltage : 100 ~ 240 V, 50/60 Hz
- Power : 25 VA
PNEUMATIC SPECIFICATIONS
- Oxygen Inlet
- Pressuried Oxygen : 0.40 Mpa, 60 L/min.
PHYSICAL
- Oxygen Inlet
- 300mm (W) x 140mm (D) x 230mm (H) Weight : 3 Kg